Analytical Instrumentation
This test is based on the latest pattern pattern of GATE for students appearing from Instrumentation - IN
Questions
In mass spectroscopy of water (H2O) peak corresponding to which mass is not possible?
- 1
- 15
- 16
- 17
- 18
Which of the following is not true about a flame photometer?
- 25-lb gauge for oxygen supply and 10-lb gauge for fuel is generally used.
- The flow rate of fuel-oxygen mixture usually varies from 2 to 10 cu. ft/h.
- An atomizer should have a tip of suitable shape to avoid a strike back.
- In flame photometry atoms in vapour state give line spectra and not band spectra.
- The usual glass or gelatin absorption filters have wide spectral bandwidths, of 35-45 nm.
Which of the following is not true about calibration curve method for determination in flame photometry?
- Calibration curves are always linear.
- In case of high total concentration of solids in the flame the burner may be fouled.
- Higher viscosity of sample should be avoided.
- If the solution is diluted excessively, errors of dilution may occur and signal to noise ration may decrease.
- For an unknown element X in a given sample, it is desirable to know that portable range of concentration of X in the sample.
Which of the following is not true about the choice of element in internal standard method for determination in flame photometry?
- The element should give spectral lines that are close to those of the analysis element.
- The reference element can be added in any concentration to those of analysis element.
- The element must be one that is not present in the samples and must be available in a high state of purity.
- The reference element must produce radiations that are perfectly selectable.
- When added the element must not alter the physical properties of the solution.
Which of the following interferences in flame photometry occurs because of the absorption of radiant energy through collision with atoms of its own kind present in ground energy level?
- Flame background emission
- Self-absorption
- Direct spectral interference
- Effect of ionization
- Effect of anions
Which of the following is not true regarding the radiation sources of infrared spectroscopy?
- The source of radiation in IR spectroscopy is ideally a black body radiator.
- Globar is a silicon carbide rod.
- Nernst filament should be externally heated to dull-red hot.
- In Globar water cooling of electrical parts are used to prevent arching.
- Nernst filament is operated in parallel with ballast resistance in constant voltage circuit.
In which of the following spectroscopy techniques does the transition occur between magnetic energy levels of paired electrons in a molecule?
- Electron spin resonance spectroscopy
- Nuclear Magnetic Resonance spectroscopy
- Raman spectroscopy
- Infrared spectroscopy
- Visible spectroscopy
Which of the following molecules cannot be analyzed with nuclear magnetic resonance spectroscopy?
- C12
- N14
- H2
- B11
- P31
Which of following is not a type of detector in IR spectroscopy?
- Pneumatic detector
- GM counter
- Bolometer
- Pyroelectric detector
- Photodiode
Which of the following statements is true regarding the sensitivity of Electron Spin Resonance (ESR)?
- ESR sensitivity increases with increasing temperature and decreasing magnetic field.
- ESR sensitivity increases with increasing temperature and increasing magnetic field.
- ESR sensitivity increases with decreasing temperature and increasing magnetic field.
- ESR sensitivity increases with decreasing temperature and decreasing magnetic field.
- The sensitivity does not depend on temperature and magnetic field.
Which of the following statements is true regarding different types of energies associated with a molecule?
- The molecule possesses vibrational energy which is associated with transitions of an electron from ground energy level to excited state level.
- When the centre of gravity does not change during to and fro motion of nuclei of the molecule, the molecule is said to posses translational energy.
- During the motion if the centre of gravity of the molecule changes the molecule which possesses vibrational energy.
- Considering an axis perpendicular to the internuclear axis and passing through the centre of gravity of the molecule, then the molecule needs electronic energy to rotate about this axis.
- Electronic Energy > Vibrational Energy > Rotational Energy > Translational Energy
Which of the following radiation detectors converts radiation energy to light energy?
- Ionisation chamber
- Scintillation counter
- Semiconductor detectors
- Proportional counter
- Geiger-Muller counter
In a scintillation counter for radioactive detection, which of the following element liquid scintillator can be used?
- C14
- Hg203
- Fe59
- Zn65
- Cd109
Which of the following is not true for a photomultiplier tube (PMT)?
- PMTs have high internal gain.
- The working of PMT is based upon photoelectric effect.
- The material of photocathode is usually a mixture of alkali metals
- The dynodes are maintained at successively more negative potential.
- The response of a PMT is linear.
Which of the following is not true regarding Beer-Lambert law?
- It defines relationship between Absorbance and Transmittance.
- Absorbance is the property of a sample while absorptivity is property of substance.
- Radiant energy will diminish in geometric or exponential progression.
- This law is true for all types of light.
- According to this law if 100% intensity is incident on first layer, then after 4th year the remaining intensity will be 6.25% (considering equal thickness layers)